R&S AFQ100A
Segment Address -
Segment Trigger
1401.3084.32
Bus
The segment trigger event is executed by remote
control using the command TRIG:SEGM:EXEC.
Remote-control command:
SOUR:TRIG:SEGM:SOUR BUS
Auto
(segment mode Single or Repeat N Times only)
The segment trigger event is executed internally
by the firmware. Segment trigger events are
ignored.
Remote-control command:
SOUR:TRIG:SEGM:SOUR AUTO
Note:
Selects the segment sequence mode of a multi segment waveform.
Note:
The Segment Address mode can be selected only by using
external segment trigger sources such as Manual, External or
Bus. With segment trigger source Auto the Segment Address
mode is not displayed.
With the aid of external signals the sequence of the segments can be
set. The signals are fed in via the BERT BNC connectors at the rear
panel of the instrument. These control signals for segment hopping
can be fed in parallel or serial.
Note:
The Segment Address mode FHOP disables the BERT
functionality and vice versa.
Auto Increment
The indices of the segments are incremented
automatically.
Remote-control command:
SOUR:WSEG:NEXT:MODE AINC
4.23
The trigger source Auto disables the segment
trigger modes Continuous and Continuous
seamless and vice versa.