Invalid block data (-161)

The command contains illegal block data.
An END message was received before the expected number of data had been
Example:
received or no numeric data element is sent after the introductory #.
Command error - sets bit 5 in the ESR register.
SCPI:

Invalid Character (-101)

The command contains an invalid sign.
A header contains an ampersand, "SOURCE&".
Example:
Command error - sets bit 5 in the ESR register.
SCPI:

Invalid separator (-103)

The command contains an impermissible sign instead of a separator.
A semicolon is missing after the first command in a command line with several
Example:
commands
":FREQ 2MHz POW -25"
Command error - sets bit 5 in the ESR register.
SCPI:

Invalid suffix (-131)

The suffix is not appropriate for this command.
nHz is not defined.
Example:
Command error - sets bit 5 in the ESR register.
SCPI:

Lists not same length (-226)

The parts of a list have different lengths. This error message is also displayed if only part of a list has
been transmitted via IEC bus. All parts of the list have to be transmitted always before it is executed.
The POWer list content is longer than the FREQuency list content, or only the
Example:
POWer content is transmitted.
Execution error - sets bit 4 in the ESR register.
SCPI: