Figure 3.138: SAS: Assign SAS Address Alias
Assign a meaningful name to each SAS address in use and click OK. The assigned names
replace the SAS address in the sample view, Search, filter, and Statistical report.
Figure 3.139: SAS:Address Alias
If you elect to save the captured sample file, the assigned SAS address names are saved
together with the result, so that when you open the sample file later, the assigned names
are retained.
Set As Default
If you want to set these SAS address aliases for sample files that will be captured later,
you can set them as default, and new samples will be opened by these default SAS
address aliases.
3.17
TxRx Vout & Preemphasis
The analyzer incorporates the ability to select TX Vout for the transmitter on each port.
Selecting TX Vout increases the output voltage swing above the nominal value, for test
and characterization purposes. This feature is also useful to compensate for line loss
when driving long cables.
324
Sierra M6‐2 SAS/SATA Protocol Analyzer User Manual
TxRx Vout & Preemphasis