Table 5-48 Spectrum Monitor Measurement Result

Value
Definition
<string 2>
Absolute peak RF frequency
[Hz]
<real 3>
Absolute peak RF level
[dBm]
<real 4>
Relative RF level [dB]
<real 5>
Relative RF level [dB]
N9360A Multi UE Tester GSM Programming Manual
Marker Measurement
TXANalyzer:RESults:SMONitor:MARKer?
This query command returns values at marker position on the
Spectrum Monitor screen with values <int 1>, <int 2>,
<real 3>, <real 4> and <real 5>.
• <int 1> (1 digit) is 0 if measurement results are within the
measurement range, or 1 if any measurement result is out of
range or the Spectrum Monitor is not measured.
The definitions and ranges of <string 2> (10 digits), <real
3> (5 letters), <real 4> (7 letters) and <real 5> (5 letters)
are shown in the table below. If any measurement result is out
of range or the Spectrum Monitor is not measured, the values
9999999999 for <int 2>, -99.9 for <real 3> and <real
5>, and 999999 for <real 4> are returned.
Range: Span 0 to 400 kHz
824200000 to 859200000,
876200000 to 915200000,
1710200000 to 1785200000,
1850200000 to 1910200000
0 to 400000
Marker Position Control
TXANalyzer:RESults:SMONitor:MARKer:POINt UP|DOWM
TXANalyzer:RESults:SMONitor:MARKer:POINt <int>
TXANalyzer:RESults:SMONitor:MARKer:POINt?
This command moves the marker UP or DOWN in steps, or by a
numeric value <real>. The allowable range is from 0 to 400000
(Hz) for the 0 to 400 kHz span, or from -100000 to 100000 (Hz)
for the ±100 kHz span.
Programming Command Reference
Range: Span ±100 kHz
824200000 to 859200000, 876200000 to
915200000, 1710200000 to 1785200000,
1850200000 to 1910200000
–100000 to 100000