Table 5-11 Power Control Level Allowable Range And Preset Value - Agilent Technologies N9360A Programming Manual

Multi ue tester, gsm programming
Hide thumbs Also See for N9360A:
Table of Contents

Advertisement

Programming Command Reference
5

Table 5-11 Power Control Level Allowable Range and Preset Value

RFORmat
GSM850
2 (+39 dBm) to 19 (+5 dBm)
GSM900
2 (+39 dBm) to 19 (+5 dBm)
DCS1800
0 (+30 dBm) to 15 (+0 dBm), or
29 (+36 dBm) to 31 (+32 dBm)
PCS1900
0 (+30 dBm) to 15 (+0 dBm), or
29 (+36 dBm) to 31 (+32 dBm)
* Values are in the order of High, Medium, and Low ranges.
5-62
Power Control Level
CONFigure:CONDition:PLEVel:HIGH <int>
CONFigure:CONDition:PLEVel:HIGH?
CONFigure:CONDition:PLEVel:MEDium <int>
CONFigure:CONDition:PLEVel:MEDium?
CONFigure:CONDition:PLEVel:LOW <int>
CONFigure:CONDition:PLEVel:LOW?
CONFigure:CONDition:PLEVel:MAN <int>
CONFigure:CONDition:PLEVel:MAN?
These commands set a value of Power Control Level (PWR
CNTL) to each of the High, Medium and Low ranges for
Automatic Test and Manual Test, for each radio system set by
the
Radio System
command described on
allowable range and preset values are as follows.
Allowable Range
This query command returns the set Power Control Level with
integer. The range of the value is same as that of the command
parameter.
page
Preset Value
*
AUTO
5, 10, 15
5, 10, 15
0, 5, 10
0, 5, 10
N9360A Multi UE Tester GSM Programming Manual
5-34. The
MANUAL
5
5
0
0

Advertisement