Spectrum Measurement - Agilent Technologies E5035A Operation Manual

Hard disk read/write test system
Table of Contents

Advertisement

Spectrum Measurement

Overview
This section describes the measurement of noise signal using a spectrum analyzer.
Measurement
1. To open the Spectrum Measurement menu, click on [Measure Spectrum] button from
Procedure
the Measure menu.
Figure 3-102
Spectrum Measurement Menu
NOTE
Before you execute this measurement, you should set an appropriate input range or execute
the Auto Configuration to set the input range automatically.
Unload
Load
Abort
2. Set the parameters, click on each button to enter a new value.
Detection Mode
RBW
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
VEE Measurement Program Operation
Unloads the head from the media. Head noise
is measured.
Loads the head on the media. Media noise is
measured.
Aborts unexpected long measurements.
Specifies the detection mode of the signal
analyzer. Select 'sampling mode' when
measuring noise. And select 'positive peak
mode' when measuring the signal. Refer to the
"hpe5022_setupSpectrum" function of the
programming manual for details.
Specifies the resolution band width.
Narrowing RBW will widen the dynamic
range. However, depending on the frequency
Measurement
157