Agilent Technologies N9360A Programming Manual page 128

Multi ue tester, gsm programming
Hide thumbs Also See for N9360A:
Table of Contents

Advertisement

Programming Command Reference
5
N O T E
5-70
Burst Timing Limit
CONFigure:CONDition:LIMit:BTIMing <real
LO>,<real HI>
CONFigure:CONDition:LIMit:BTIMing?
This command specifies the low and high limits for Burst
Timing tests. The allowable range is from -9.9 to 9.9 bits in
0.1 steps. At preset, these are set to -1.0 and 1.0 bits.
This query command returns the set low and high limits for
Burst Timing test with real separated by a comma. The range is
same as that of the command parameter.
Phase Error Limit
CONFigure:CONDition:LIMit:PHASe:RMS <real HI>
CONFigure:CONDition:LIMit:PHASe:RMS?
CONFigure:CONDition:LIMit:PHASe:PEAK <real HI>
CONFigure:CONDition:LIMit:PHASe:PEAK?
These commands set the high limit for Phase Error (RMS) and a
parit of the low and high limits for Phase Error (Peak) tests. The
allowable range is from 0.0 to 99.9 (°) in 0.1 steps. At preset,
these are set to 5.0 (°) for RMS and 20.0 (°) for Peak.
These query commands return the set high limit for RMS and a
parit of the low and high limits for Peak Phase Error test with
real. The range is same as that of the command parameter.
This command is invalid when EGPRS 8-PSK is selected.
N9360A Multi UE Tester GSM Programming Manual

Advertisement

Table of Contents
loading