Figure 4-7 Status Byte Register; Table 4-4 Status Byte Register - Agilent Technologies N9360A Programming Manual

Multi ue tester, gsm programming
Hide thumbs Also See for N9360A:
Table of Contents

Advertisement

Programming Command Guideline
4

Table 4-4 Status Byte Register

Bit
Description
0
0: Measurement data is not ready in the output queue.
1: Measurement data is ready in the output queue.
1
0: Measurement is completed.
1: Measurement is not completed.
2, 3
These bits are always set to 0.
4
0: Measurement data is not ready in the output queue.
1: Measurement data is ready in the output queue.
5
0: Standard Event Status Register Masked by Standard Event Status Enable Register is not set to 1.
1: Standard Event Status Register Masked by Standard Event Status Enable Register is set to 1.
6, 7
These bits are always set to 0.
4-20
Status Byte Register
The Status Byte Register of the Tester uses the bit 0
Measurement Data Ready Bit, bit 1 Measuring Status Bit, bit 4
Message Available and bit 5 Standard Event Status Summary Bit as
shown in
Figure
4-7.

Figure 4-7 Status Byte Register

The Status Byte Register contains the following information:
To query the Status Byte Register, send the command *STB?.
The response is the decimal sum of the bits which are set to
For example, if the bit 5 and bit 4 are set to
of these 2 bits is 48 (32 plus 16) see
value 48 is returned.
1
Figure
4-5. So the decimal
N9360A Multi UE Tester GSM Programming Manual
.
1
, the decimal sum

Advertisement

Table of Contents
loading

This manual is also suitable for:

Gs8210

Table of Contents