N9360A Multi UE Tester GSM Programming Manual
Marker Position Control
TESTs:MANual:RESults:SMONitor:MARKer:POINt
UP|DOWN
TESTs:MANual:RESults:SMONitor:MARKer:POINt <int>
TESTs:MANual:RESults:SMONitor:MARKer:POINt?
This command moves the marker position UP or DOWN in steps,
or by <real> ranging from 0 to 400000 in Hz.
Measurement Result
TESTs:MANual:RESults:SMONitor?
This query command returns measurement result values <int
1>, <string 2>, <real 3>, <real 4> and <real 5>. Each
value is separated by a comma.
• <int 1> (1 digit) is 0 if measurement results are within the
measurement range, or 1 if any measurement result is out of
range or the Spectrum monitor is not measured.
The definitions and ranges of <string 2> (10 digits),
<real 3> (5 letters), <real 4> (5 letters), and <real 5>
(5 letters) are shown in the table below. If any measurement
result is out of range or the Spectrum monitor is not
measured, the values 9999999999 for <string 2> and
-99.9 for <real 3>, <real 4> and <real 5>, are
returned.

Table 5-41 Spectrum Monitor Measurement Result

Value
Definition
<string 2>
Absolute peak RF frequency [Hz] 880200000 to 915200000,
<real 3>
Absolute peak RF level [dBm]
<real 4>
Relative RF level [Hz]
<real 5>
Relative RF level [dB]
Programming Command Reference
Range
1710200000 to 1785200000,
1850200000 to 1901200000
–99.9 to 99.9
–99.9 to 0.0 (at +200 kHz)
–99.9 to 0.0 (at +250 kHz)
5
5-163