Command Reference for Phase/Frequency Error Measurement Screen

N9360A Multi UE Tester GSM Programming Manual
The PHASe commands are used to set or get the controls and
parameters associated with the Phase Error and Frequency
Error measurements and the pass or fail tests.
Phase Error Marker Measurement
TESTs:MANual:RESults:PHASe:MARKer?
This query command returns values at marker position on the
Phase Error graph screen with values <int 1>, <int 2>,
<real 3>, and <real 4>. Each value is separated by comma.
• <int 1> (1 digit) is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or the Phase Error is not measured.
• <int 2> (1 digit) is the test result for the 0 (pass) or 1 (fail).
• <real 3> (up to 6 letters) is the marker position ranging
from 0.00 to 147.00 in bits.
• <real 4> (up to 5 letters) is the measured Phase error at
marker position. If measurement result is out of range or the
Phase Error is not measured, the value 99.9 is returned.
Phase Error Marker Position Control
TESTs:MANual:RESults:PHASe:MARKer:POINt UP|DOWN
TESTs:MANual:RESults:PHASe:MARKer:POINt <real>
TESTs:MANual:RESults:PHASe:MARKer:POINt?
This command moves the marker position Up or Down in steps,
or by <real> bits ranging from 0.00 to 147.00 in bits.
Programming Command Reference
5
5-153