Programming Command Reference
5
5-4
Status Byte Register Query
*STB?
This query command returns a value of the Status Byte Register
(STBR). The range of a value is from 0 to 255. Information
contained in the Status Byte Register is as follows:
bit 0 (1): Measurement Data Ready
bit 1 (2): Measuring Status
bit 2 (4): Unused
bit 3 (8): Unused
bit 4 (16): Message Available
bit 5 (32): Standard Event Status Summary
bit 6 (64): Unused
bit 7 (128): Unused
For detailed information, refer to
page
4-20.
Instrument Option Query
*OPT?
This query command returns a value of the option numbers
installed in the Tester.
Selftest Result Query
*TST?
This query command returns the result of self diagnosis with
integer value of 0 (Normal) or 1 (Abnormal).
Status Byte Register
N9360A Multi UE Tester GSM Programming Manual