Table 5-33 Power Control Level Allowable Range And Preset Value - Agilent Technologies N9360A Programming Manual

Multi ue tester, gsm programming
Hide thumbs Also See for N9360A:
Table of Contents

Advertisement

Table 5-33 Power Control Level Allowable Range and Preset Value

RFORmat
Allowable Range
GSM850
2 (+39 dBm) to 19 (+5 dBm)
GSM900
2 (+39 dBm) to 19 (+5 dBm)
DCS1800
0 (+30 dBm) to 15 (+0 dBm), or 29 (+36 dBm) to 31 (+32 dBm)
PCS1900
0 (+30 dBm) to 15 (+0 dBm), or 30 (+33 dBm) to 31 (+32 dBm)
N9360A Multi UE Tester GSM Programming Manual
This query command returns the set Power Control Level with
integer.
The range of the value is same as that of the command
parameter.
BS Level
TESTs:MANual:AMPLitude <real>
TESTs:MANual:AMPLitude?
This command sets the amplitude (BS Level) of output signal
from the Tester. The allowable range is from -110.0 to -50.0
(dBm) in 0.1 steps. At preset, this is set to -50.0 (dBm).
This query command returns the set amplitude of output signal
with real. The range of the value is same as that of the command
parameter.
Slot Configuration
TESTs:MANual:SCONfigure 1
TESTs:MANual:SCONfigure?
This command selects the slot combination for downlink and
uplink, respectively, on the [Stand-by] screen for the
GPRS/EGPRS mode, in conjunction with the Slot Order
Measured setting. The combinations of the slot configuration
and the slot order measured are as follows:
Programming Command Reference
Preset Value
5 (+33 dBm)
5 (+33 dBm)
0 (+30 dBm)
0 (+30 dBm)
×
×
×
×
1|2
1|2
2|3
1|3
5
×
×
2|4
1
5-123