Agilent Technologies N9360A Programming Manual page 233

Multi ue tester, gsm programming
Hide thumbs Also See for N9360A:
Table of Contents

Advertisement

N9360A Multi UE Tester GSM Programming Manual
Power Ramp Flatness
TXANalyzer:RESults:RAMP:FLATness?
This query command returns the maximum value and the
minimum value in the burst on period.
• <int1> is 0 if measurement result is within the
measurement range, or 1 if measurement result is out of
range or the Access Probe Power is not measured.
• <real2> is maximum value (dB). If measurement result is
out of range or the Access Probe Power is not measured, the
value -99.9 is returned.
• <real3> is minimum value (dB). If measurement result is
out of range or the Access Probe Power is not measured, the
value -99.9 is returned.
Power Ramp Trace Start
TXANalyzer:RESults:RAMP:TRACe:STARt <real>
TXANalyzer:RESults:RAMP:TRACe:STARt?
This command sets the start point of trace data in Power Ramp
for the query command.
The range of a value is from -9.250 to 156.250 Bits in GMSK.
The step is 0.125 Bits step and the number of data is 1325 in
GMSK.
The range of a value is from -27.750 to 468.750 Bits in 8PSK.
The step is 0.375 Bits step and the number of data is 1325 in
8PSK.
The set value by this command is a common value to TCH1 to
TCH6 Manual Test and TX Analyzer.
Programming Command Reference
5
5-175