Advertisement

Programming Command Reference
5
N O T E
N O T E
5-74
BLER (GPRS Mode, Test Mode ACK) Limit
CONFigure:CONDition:LIMit:BLER <real HI>
CONFigure:CONDition:LIMit:BLER?
This command sets the high limit for Block Error Ratio (BLER)
test if the GPRS/EGPRS mode is selected. The allowable range is
from 0.00 to 99.99 (%) in 0.01 steps. At preset, this is set to
10.00 (%).
This query command returns the set high limit for BLER test
with real. The range is same as that of the command parameter.
If the GSM mode is set to "GPRS" or "EGPRS", Sensitivity in BLER is
measured in Manual Test.
EVM
CONFigure:CONDition:LIMit:EVM:RMS <real HI>
CONFigure:CONDition:LIMit:EVM:RMS?
CONFigure:CONDition:LIMit:EVM:PEAK <real HI>
CONFigure:CONDition:LIMit:EVM:PEAK?
CONFigure:CONDition:LIMit:EVM:Ninetyfive <real
HI>
CONFigure:CONDition:LIMit:EVM:Ninetyfive?
These commands set high limit for EVM Test. The allowable
range is from 0.00 to 99.99 (%) in 0.01 steps. The preset value of
RMS, PEAK, and Ninetyfive is set to 9.00, 30.00 and 15.00
respectively.
This command is valid when the GSM mode is set to EGPRS.
N9360A Multi UE Tester GSM Programming Manual