Agilent Technologies N9360A Programming Manual page 218

Multi ue tester, gsm programming
Hide thumbs Also See for N9360A:
Table of Contents

Advertisement

Programming Command Reference
5
5-160
ORFS Switching
TESTs:MANual:ORFS:SW:STATe ON|OFF|0|1
TESTs:MANual:ORFS:SW:STATe?
This command sets whether to execute each measurement item
of the ORFS Switching. The parameter is ON (1) or OFF (0). At
preset, this is set to ON.
Measurement Count
TESTs:MANual:ORFSCount <int>
TESTs:MANual:ORFSCount?
This command sets ORFS Measurement count in EGPRS mode.
The parameter is ranging from 1 to 200. At preset, this is set to
1.
ORFS Modulation Measurement
TESTs:MANual:RESults:ORFS:MOD?
This query command returns the ORFS Modulation
measurement result values <int 1>, <int 2>, <real 3> and
<real 4>. Each value is separated by a comma.
• <int 1> (1 digit) is 0 if measurement results are within the
measurement range, or 1 if any measurement result is out of
range or the ORFS Modulation is not measured.
• <int 2> (1digit) is the test result for the 0 (pass) or 1 (fail).
• <real 3> is the Lower value and <real 4> is the Upper
value of the measurement results. If the measurement result
is out of range or the ORFS Modulation is not measured, the
value 99.99 is returned.
N9360A Multi UE Tester GSM Programming Manual