Command Reference for Configuration: Test Condition (Loss) Screen

5-78
Attenuation Control
CONFigure:LOSS ON|OFF|1|0
CONFigure:LOSS?
This command sets the attenuation control function to ON (1) or
OFF (0). If this is set to ON, the attenuation control function for
each band is activated and is ready to be used in the RF input
path and RF output path. At preset, this is set to 0.
This query command returns the attenuation function setting
by integer value of 1 or 0.
Attenuation Level Setting for Input Signal
CONFigure:LOSS:RFIN <real 1>,<real 2>,<real
3>,<real 4>
CONFigure:LOSS:RFIN?
This command specifies the losses between a mobile phone and
the Tester caused by the RF Input path for radio system of
GSM850, GSM900, DCS1800, and PCS1900, respectively. The
allowable range is from 0.0 to 99.9 (dB) in 0.1 steps. At preset,
it is set to 0.0 (dB).
This query command returns the loss setting with real value.
The range of the values is same as that of the command
parameter.
Attenuation Level Setting for Output Signal
CONFigure:LOSS:RFOUt <real 1>,<real 2>,<real
3>,<real 4>
CONFigure:LOSS:RFOUt?
This command specifies the insertion losses between a mobile
phone and the Tester caused by the RF Output path for radio
system of GSM850, GSM900, DCS1800, and PCS1900,
respectively. The allowable range is from 0.0 to 99.9 (dB) in
0.1 steps. At preset, it is set to 0.0 (dB).
This query command returns the loss setting between a mobile
phone and the Tester with real value. The range of the values is
same as that of command parameter.