Programming Command Reference
5
5-192
Frequency Offset
RFGenerator:OFFSet <real>
RFGenerator:OFFSet?
This command selects the frequency offset. The allowable range
is from -100.0 to 100.0 (kHz) in 0.1 steps. At preset, this is
set to 0.0 (kHz).
This query command returns the set frequency offset with real.
The range of the value is same as that of the command
parameter.
Modulation
RFGenerator:MODulation <<
OFF|BCCH|PN9|PN15|ALL0|ALL1|AM|PSK9|PSK15
RFGenerator:MODulation?
This command selects a type of Modulation. The parameter is
OFF, BCCH, PN9, PN15, ALL0, ALL1, AM, PSK9 and PSK15. At
preset, this is set to OFF.
GSMK
• If OFF is selected, the output signal is a continuous wave
(CW).
• GMSK:BCCH
GMSK modulation with BCCH data.
• GMSK:PN9
GMSK modulation with PN9 data.
• GMSK:PN15
GMSK modulation with PN15 data.
• GMSK:All0
GMSK modulation with all-zero data.
• GMSK:All1
GMSK modulation with all-one data.
• AM
Applies AM modulation to the carrier.