Table 11 Equipment Required To Verify Trigger Sensitivity - Agilent Technologies 6000 Series Service Manual

Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

2
Testing Performance
52
Test limits for:
• Internal trigger on all models:
< 10 mV/div: greater of 1 div or 5 mV
>=10 mV/div: 0.6 div
• External trigger on all 2-channel models (DSO/MSO6xx2A):
Trigger range: 1.0V
DC to 100 MHz: < 100 mV
>100 MHz to max bandwidth: < 200 mV
Trigger range: 8.0V
DC to 100 MHz: < 250 mV
>100 MHz to max bandwidth: < 500 mV
• External trigger on 4-channel models:
4-channel 1 GHz, 500 MHz, and 300 MHz models:
(MSO/DSO6104A/6054A/6034A, DSO6104L/6054L)
DC to 500 MHz: < 500 mV
4-channel 100 MHz models:
(MSO/DSO6014A, DSO6014L)
DC to 100 MHz: < 500 mV
Table 11
Equipment Required to Verify Trigger Sensitivity
Equipment
Critical Specifications
Signal
25-MHz, 100-MHz, 300-MHz,
Generator
500-MHz, and 1 GHz sine
waves
Power splitter Outputs differ < 0.15 dB
Cable
BNC, Qty 3
Adapter
N (m) to BNC (f), Qty 3
Feedthrough
50Ω BNC (f) to BNC (m) Qty 2
pp
pp
pp
pp
pp
pp
pp
Recommended Model/Part
Agilent N5181A, or E4400B, or
8648A
Agilent 11667A
Agilent 10503A
Agilent 1250-0780
Agilent 0960-0301