Replace A Column - Single Column Gcs - Agilent Technologies Intuvo 9000 Series Maintaining

Gas chromatograph
Table of Contents

Advertisement

2
Maintaining Columns and Bus Components
Replace a Column – Single Column GCs
WA RNING
28
This procedure applies to GCs using a single column. A single
column is always installed in the right-side position.
The inlet, detector, bus components, and column can be hot
enough to cause burns. Cool heated zones to a safe handling
temperature before continuing.
Launch the GC maintenance wizard: Maintenance > Columns >
1
Perform Maintenance > Replace Column > Start Maintenance. The
wizard will walk through the replacement procedure. These
steps are repeated below.
2
Open the column oven.
Figure 8
GC shown configured for a single column (right side)
a
Open the GC front door.
b
Remove the bus door.
c
Lower the oven door.
Right-side column
SmartID key
Maintaining Your GC