Next Pk Left; Marker Delta; Min Search - Agilent Technologies X Series Manual

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Notes
State Saved

Next Pk Left

Moves the selected marker to the nearest peak left of the current marker.
If the selected marker was off, then it is turned on as a normal marker and a peak search is performed.
Key Path
Remote Command
Example
State Saved

Marker Delta

Performs the same function as the Delta 1-of-N selection key in the Marker menu. Basically, this sets the
control mode for the selected marker to Delta mode. See the Marker section for details of delta markers.
The key is duplicated here in the Peak Search Menu to allow you to conveniently perform a peak search
and change the marker's control mode to Delta without having to access two separate menus.
Key Path
Notes
Initial S/W Revision

Min Search

Moves the selected marker to the minimum y-axis value on the current trace.
Key Path
Mode
Remote Command
Example
Notes
Initial S/W Revision
Sending this command selects the subopcoded marker
Not part of saved state
Peak Search
:CALCulate:SPECtrum:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:
MAXimum:LEFT
CALC:SPEC:MARK2:MAX:LEFT
Selects marker 2 and moves it to the next peak to the left of the current marker
position.
Not part of saved state
Peak Search or Marker
Whenever the selected marker is in Delta mode and you are in the Peak Search
menu, the Marker Delta key should be highlighted and the active function for
setting its delta value turned on.
Prior to A.02.00
Peak Search
BASIC
:CALCulate:SPECtrum:MARKer[1]|2|3|4|5|6|7|8|9|10|11|12:
MINimum
CALC:SPEC:MARK:MIN
You must be in the IQ Analyzer (Basic) mode to use this command. Use
INSTrument:SELect to set the mode.
Prior to A.02.00
Complex Spectrum Measurement
Peak Search