Performance Test and Function Test

Performance Test (for Option 010 Probe Station Connection Kit)

Performance Test (for Option 010 Probe Station Connection
Kit)

Introduction

This section provides the information on the performance test setups and procedures for the
E4991A Option 010 Probe Station Connection Kit. To execute the performance tests for
the Option 010, both the test head for Option 010 and the E4991A are required.

Required Test

The E4991A with Option 010 requires the following performance tests with the test setups
for Option 010 to be done in addition to the performance tests for standard E4991A.
Frequency Accuracy Test
Power Level Accuracy Test
Impedance Measurement Accuracy Test
DC Bias Level Accuracy Test (Opt. 001)

Specification

Except for the Power Level Accuracy, the E4991A Option 010 specifications are same as
the standard E4991A specifications. Therefore, the Frequency Accuracy, Impedance
Measurement Accuracy and DC Bias Level Accuracy of the E4991A with Option 010 are
tested to the same test limits as the standard E4991A. The Power Level Accuracy
specification of the Option 010 is different from that of the standard E4991A as follows:
Power Level Accuracy:
48
2 dB @  1 GHz (without Opt. 010)
5.5 dB @  1 GHz (with Opt. 010)
3 dB @ > 1 GHz (without Opt. 010)
7.6 dB @ > 1 GHz (with Opt. 010)
Chapter 2