Selecting BIN range mode

You can select whether the BIN range should be inside or outside of the set limit values
(upper and lower limit values) or whether the BIN range should cover the entire range
regardless of the limit values.
Step 1. Click with the mouse in the cell corresponding to the column of the desired sorting
condition (one of conditions 1 to 4) on the In/Out row of the BIN to be set.
Step 2. Click the down arrow () at the right edge of the cell.
Step 3. Click the desired BIN range mode from the drop-down list box to select it.
Bin range mode
In
Out
All
Figure 8-11
Selecting BIN range mode
Chapter 8
Sorting According to Measurement Results
Setting BIN Range
Definition
Inside. This mode ensures that the BIN
range is inside the set limit values (upper
and lower limit values).
Outside. This mode ensures that the BIN
range is outside the set limit values (upper
and lower limit values). Namely, the BIN
is used when the result is equal to or
greater than the upper limit value or equal
to or less than the lower limit value.
The entire range. In this mode the BIN
range covers the entire range regardless of
the set limit values (upper and lower limit
values). Namely, the system assumes that
any measurement result is within the set
range.