How To Change Measurement Parameters - Agilent Technologies Twelfth 4287A RF LCR Operation Manual

Table of Contents

Advertisement

Learning Basic Operations
Display of DUT Connection and Measurement Result
Step 5. Press the rotary knob (
(Figure 3-29).
Figure 3-29
Measurement point changed to measurement point number 4

How to change measurement parameters

Set which of the four measurement parameters is to appear on the display by following
these steps.
Step 1. Press the
The measurement parameter selection menu appears in the softkey area on the right side of
the display.
Step 2. Turn the rotary knob (
PRMTRS
Step 3. Press the rotary knob (
right side of the display.
Step 4. Turn the rotary knob (
corresponding to the measurement point number whose measurement parameters should be
changed (
Step 5. Press the rotary knob (
the display.
Step 6. Turn the rotary knob (
corresponding to the desired measurement parameter (
Step 7. Press the rotary knob (
70
) to accept the selection of the measurement point number
[Prmtr]
key on the measurement block (MEASUREMENT) of the front panel.
) to move the cursor in the softkey menu to the
key.
) to display the SETUP MEAS PRMTRS softkey menu on the
) to move the cursor in the softkey menu to the softkey
in this case).
PRMTR-1 [ ]
) to display the PRMTR-1 [ ] softkey menu on the right side of
) to move the cursor in the softkey menu to the softkey
) to accept the selection of the measurement parameter (Figure
SETUP MEAS
in this case).
Ls
Chapter 3