Bin sorting setup display

When no dialog box is shown, the BIN sorting setup display (Figure 4-10) can be displayed
by pressing the
[Setup View]
key the required number of times.
Figure 4-10
Bin sorting setup display
1. Comparator (bin sorting) setup tab
When the "COMPARATOR" tab is displayed in front of the other three tabs, the screen
shows the bin sorting setup display. If another setup display appears, click this tab with the
mouse to get the bin sorting setup display. (The same operation can be done with the
"View" menu in the menu bar.)
2. Table number
Displays the currently selected table number.
3. DC resistance (Rdc) upper and lower limit values
You can set the upper and lower limit values for the DC resistance (Rdc) measurement
result.
4. BIN number, BIN ON/OFF
Use this button to set the bin number and On/Off state. You can set the conditions for
sorting parts into a maximum of 13 bins. If you turn off a bin, the bin is regarded as
non-existent.