Specifications and Supplemental Performance Characteristics
Basic Measurement Characteristics
Measurement parameters
Impedance parameters
Measurement range
Measurement range
Source characteristics
Frequency
Range
Resolution
Uncertainty
Oscillator Level
Range
Cable length: 1 m
Power (when 50 Ω LOAD is
connected to test port)
Current (when SHORT is
connected to test port)
Voltage (when OPEN is
connected to test port)
Cable length: 2 m
Power
Chapter 12
Basic Measurement Characteristics
|Z|, |Y|, Ls, Lp, Cs, Cp, Rs, Rp, X, G, B, D, Q, θz [°],
θz [rad], θy [°], θy [rad]
(A maximum of four parameters can be displayed at one
time.)
200 mΩ to 3 kΩ
Frequency = 1 MHz
Averaging factor = 8
Oscillator level ≥ −33 dBm
Measurement uncertainty ≤ ±10 %
Calibration is performed within 23 °C ±5 °C
Measurement is performed within ±5 °C of calibration
temperature)
1 MHz to 3 GHz
100 kHz
±10 ppm (23 ±5°C)
±20 ppm (5°C to 40°C)
− 40 dBm to 1 dBm (Frequency ≤ 1 GHz)
− 40 dBm to 0 dBm (Frequency > 1GHz
0.0894 mArms to 10 mArms (Frequency ≤ 1 GHz)
0.0894 mArms to 8.94 mArms (Frequency > 1 GHz
4.47 mVrms to 502 mVrms (Frequency ≤ 1 GHz)
4.47 mVrms to 447 mVrms (Frequency > 1 GHz
(when Option 002 is used)
Subtract the following attenuation from the power (setting
value) at 1 m cable length:
Attenuation [dB] = 0.37 ×
F
(F: Frequency [GHz])
*1
)