Agilent Technologies Twelfth 4287A RF LCR Operation Manual page 168

Table of Contents

Advertisement

Executing Measurements and Displaying Results
Turning ON/OFF Update the LCD Display View
Figure 7-6
View when the update of the screen display is turned off
Table 7-2
Time of each task (typical values)
Conditions:
Measurement instruments:
NOTE
When performing measurements with the screen display set to ON, there are several
factors that need to be considered. Therefore, make the most appropriate selection for the
instrument settings by taking the following points into account.
1. When the screen display is set to perform updates (UPDATE:ON) and a GPIB
168
Measurement sequence
Setup time of READ?
Time from TRIG-EOM
Data acquisition time
Display time + internal
processing time
Ls-Q measurement, single point measurement (LIST measurement), Rdc measurement,
comparator OFF, Measurement signal level monitor ON.
Host computer: i486DX4, 100 MHz / 82335B GPIB Card
Data acquisition time is transfer time using ASCII format.
command is sent to the instrument, the instrument will perform processes to update its
DISP:OFF
UPDATE:ON
5.3 ms
33.0 ms
Time from T2 to T5 (11.1 ms max)
3.2 ms
3.2 ms
0 ms
10.3 ms
DISP:ON
UPDATE:OFF
5.6 ms
3.2 ms
10.3 ms
Chapter 7