Sorting According to Measurement Results
Example of Settings for BIN Sorting
NOTE
Instead of setting the upper limit to infinite, 1G is entered as a sufficiently large value.
The In/Out cell is left unchanged in the default setting "In" (the BIN range is between the
upper limit and lower limit).
Step 5. Set the lower limit measurement value for sorting into BIN 1 for Condition 2 by following
this procedure.
a. Click the BIN 1 LowerLimit cell in the Condition 2 column to select the cell.
b. Press the [1][0][0] keys in the entry block on the front panel (Figure 8-28) and then
Figure 8-28
Setting the lower limit value for Condition 2 of BIN 1
Step 6. Set the BIN range mode of Condition 2 for sorting into BIN 1 to "In".
200
press the
key to enter 100 as the lower limit value for Condition 2 of BIN 1.
[×1]
Chapter 8