Turning On BIN Sorting and Displaying the Result

When BIN sorting setup is completed, follow these steps to turn on the BIN sorting
function.
Step 1. Press the
System softkey menu.
Step 2. Press the
execution (the setting toggles with each press of the key).
Softkey Label
COMPARATOR [ON]
COMPARATOR [OFF]
When BIN sorting is turned on, the BIN sorting result is displayed for each measurement in
the BIN sorting result area in the upper left area of the display (Table 8-15).
Figure 8-15
BIN Sort Result Display
NOTE
The figure displayed following "BIN =" is the BIN number as the BIN sort result.
When the measurement result is not sorted in any of the BINs, "BIN = 0" is displayed.
When Overload or Rdc limit deviation occurred, "BIN = 14" is displayed.
When the measurement result is sorted in "Good Bins" defined by "Setting Faulty BIN
(OUT OF GOOD BINS)" on page 187, the BIN sort result is displayed in white; when the
measurement result is sorted in "Faulty Bins," the BIN sort result is displayed in red.
Chapter 8
Turning On BIN Sorting and Displaying the Result
key in the Instrument State Block on the front panel to display the
[System]
key in the softkey menu to turn ON/OFF the BIN sorting
COMPARATOR [ ]
Sorting According to Measurement Results
Display Status
ON (BIN sorting is executed)
OFF (BIN sorting is not executed)