Agilent Technologies Twelfth 4287A RF LCR Operation Manual page 125

Table of Contents

Advertisement

Figure 6-8
How to connect short standard
Step 2. Click the
Step 3. Turn the short standard counterclockwise viewed from the top and remove the short
standard from the 7-mm terminal.
7. How to measure load calibration data
Step 1. Turn the connector nut outside the load standard (50 Ω) counterclockwise viewed from the
side where the connector is mounted and completely retract the internal connector sleeve
(Figure 6-9).
Figure 6-9
How to retract load standard connector sleeve
Step 2. Firmly connect the load standard to the 7-mm terminal on the test head by turning the
connector nut clockwise as viewed from the top (Figure 6-10).
Chapter 6
Calibration and Compensation Procedures Using Dedicated Test Fixture
button in the calibration window and measure the short calibration data.
SHORT
Calibration and Compensation
125