Executing Open Calibration - Agilent Technologies Twelfth 4287A RF LCR Operation Manual

Table of Contents

Advertisement

Figure 3-14
How to connect short standard
NOTE
After turning the connector nut of the short standard and attaching it to the 7-mm terminal,
it is not necessary to turn the connector nut of the 7-mm terminal on the test head side to fix
it in position. This also applies to the installations of the open standard, low-loss capacitor
and load standard explained in this section.
Step 11. Press the
cursor to the
Step 12. Press the
in the Calibration progress indicator in the calibration window indicates the progress of the
short calibration data measurement. Upon completion of short calibration data
measurement, the screen changes to the next window (Figure 3-15).

Executing open calibration

Figure 3-15
Open calibration window
Step 1. Turn the outside of the short standard counterclockwise and remove it from the 7-mm
terminal.
Chapter 3
key or
key on the front panel the required number of times to move the
button.
Next>
key on the front panel to start measurement of short calibration data. The bar
Learning Basic Operations
Calibration and Compensation
61