Calibration and Compensation
Calibration and Compensation Procedures Using Dedicated Test Fixture
Figure 6-6
How to extend the 7-mm connector sleeve
Step 2. Firmly connect the open standard (0 S) to the 7-mm terminal on the test head by turning it
clockwise as viewed from the top (Figure 6-7).
Figure 6-7
How to connect open standard
Step 3. Click the
Step 4. Turn the open standard counterclockwise as viewed from the top and remove it from the
7-mm terminal.
6. How to measure short calibration data
Step 1. Firmly connect the short standard (0 Ω) to the 7-mm terminal on the test head by turning it
clockwise as viewed from the top (Figure 6-8).
124
OPEN
button in the calibration window and measure the open calibration data.
Chapter 6