Calibration Kit Setup Display - Agilent Technologies Twelfth 4287A RF LCR Operation Manual

Table of Contents

Advertisement

Function Overview
Names and Functions of Display Areas on the LCD Display

Calibration kit setup display

When no dialog box is shown, the calibration kit setup display (Figure 4-8) can be
displayed by pressing the
[Setup View]
key the required number of times.
Figure 4-8
Calibration kit setup display
1. Calibration kit setup tab
When the "CAL KIT" tab is displayed in front of the other three tabs, the screen shows the
calibration kit setup display. If another setup display appears, click this tab with the mouse
to get the calibration kit setup display. (The same operation can be done with the "View"
menu in the menu bar.)
2. Calibration kit selection
You can choose whether to use the values predetermined for the 7-mm calibration kit
(7-mm standard) or user-defined values as a calibration kit definition. To use the Agilent
Technologies 16195B 7-mm calibration kit, select "7 mm Standard."
Before selecting "7 mm Standard," all calibration kit values should have been entered; the
user is not allowed to change the definition.
If you select "User Defined," you can freely enter the values for the calibration kit
definition. If you check (√) "Fixed Model," enter the definition for measurement point
number 1, which is then automatically applied to the other measurement point numbers.
3. Measurement point number/measurement frequency
Displays the measurement points (measurement point number and measurement
frequency) in the selected measurement point table.