Function Overview
Names and Functions of Display Areas on the LCD Display
4. Open standard value
Enter the open standard value.
5. Short standard value
Enter the short standard value.
6. Load standard value
Enter the load standard value.
7. Circuit model
Select one of the three circuit models (Rs-Ls, Ls-Q, or Cp-D) for the load parameter.
8. Offset delay
Depending on the calibration kit used, offset delay may have to be set. In this case, enter
the offset delay values for open standard, short standard, and load standard.
9. DC standard value
Enter the calibration kit definition for 4287A DC resistance (Rdc) measurement.