Connection of Test Fixture

Connect the test fixture as follows:
Figure 3-22
Test fixture connection window
Step 1. Turn the connector nut of the low-loss capacitor counterclockwise and remove it from the
7-mm terminal.
Step 2. Turn the connector nut of the 7-mm connector counterclockwise to retract the connector
sleeve completely inside (Figure 3-23).
Figure 3-23
Retracting the 7-mm connector connector sleeve
Step 3. Fit the tips of the two mounting posts on the test fixture stand to the two holes of the test
fixture and fit the 7-mm connector on the test fixture stand to the 7-mm connector on the
test fixture. Turn the connector nut of the 7-mm connector on the test fixture stand
counterclockwise, as viewed from above, and firmly lock it to the test fixture's 7-mm
connector.
Chapter 3
Learning Basic Operations
Calibration and Compensation
65