Removing center conductor collet
Follow these steps to remove the center conductor collet of the 7-mm connector.
Step 1. Turn the outer part of the 7-mm connector clockwise viewed from above to completely
expose the connector sleeve.
Step 2. Pull up the handle of the collet replacement cool (Figure A below) and, while keeping
pulling the handle, insert the replacement tool slowly until it touches the end of the
connector (Figure B below).
Step 3. After releasing the handle (Figure A below), pull up the collet removal tool (Figure B
below) to remove the old collet.
Appendix G
Information on Maintenance
Cleaning this Instrument
331