NOTE
If you want to use the open state of the test fixture instead of connecting the open standard,
set the open state according to the Operation Manual of the test fixture.
Step 2. Click the
6. How to measure short calibration data
Measure the short calibration data as follows:
Step 1. Connect the short standard to the DUT connection terminal of the test fixture.
NOTE
If you want to use the short state of the test fixture instead of connecting the short standard,
set the short state according to the Operation Manual of the test fixture.
Step 2. Click the
7. How to measure load calibration data
Measure the load calibration data as follows:
Step 1. Connect the load standard to the DUT connection terminal of the test fixture.
Step 2. Click the
8. How to measure DC open/short/load calibration data
When you want to use the dedicated working standard to measure DC resistance (Rdc),
follow these steps to execute additional DC open/short/load calibration.
Step 1. Connect the DC open standard to the DUT connection terminal of the test fixture.
Step 2. Click the
data.
Step 3. Connect the DC short standard to the DUT connection terminal of the test fixture.
Step 4. Click the
calibration data.
Step 5. Connect the load standard to the DUT connection terminal of the test fixture.
Step 6. Click the
data.
9. Specifying creation of automatic setting file for when power is turned ON
The setting conditions of the 4287A, including calibration data and compensation data, can
be automatically stored for reuse in a file named "autorec.sta" in the internal nonvolatile
memory. To use this function, click the
calibration so that the check mark (
NOTE
When the 4287A is turned ON, it searches the internal nonvolatile memory for a file called
"autorec.sta," and if it is found, the setting conditions (including calibration and
compensation data) stored in the file are automatically recalled.
Chapter 6
Calibration and Compensation Procedures Using Dedicated Test Fixture
OPEN
button in the calibration window and measure the open calibration data.
SHORT
button in the calibration window and measure the short calibration data.
button in the calibration window and measure the load calibration data.
LOAD
button in the calibration window and measure the DC open calibration
DC OPEN
button in the calibration window and measure the DC short
DC SHORT
DC LOAD
button in the calibration window and measure the DC load calibration
Calibration and Compensation
check box upon completion of
Auto Recall
) appears.
133