Agilent Technologies Twelfth 4287A RF LCR Operation Manual page 127

Table of Contents

Advertisement

Figure 6-11
How to connect low-loss capacitor
Step 2. Click the
capacitor data.
Step 3. Turn the connector nut outside the low-loss capacitor counterclockwise as viewed from the
top and remove it from the 7-mm terminal.
10. Specifying creation of automatic setting file for when power is turned ON
The setting conditions of the 4287A, including calibration data and compensation data, can
be automatically stored for reuse in a file named "autorec.sta" in the internal nonvolatile
memory. To use this function, click the Auto Recall check box upon completion of
calibration so that the check mark (√) appears.
NOTE
When the 4287A is turned ON, it searches the internal nonvolatile memory for a file called
"autorec.sta," and if it is found, the setting conditions (including calibration and
compensation data) stored in the file are automatically recalled.
As described in "20. Specifying creation of automatic setting file for when power is turned
ON" on page 130, it is also possible to automatically store the setting conditions, including
the compensation data, in a file named "autorec.sta" upon completion of compensation.
When this is done, the calibration data executed immediately before compensation are
simultaneously saved. Accordingly, if both calibration and compensation are to be
executed, it is not necessary to click the
11. How to exit calibration
To exit calibration, click the
12. How to mount dedicated test fixture
Mount the dedicated test fixture on the 7-mm terminal of the test head. For the mounting
procedure, see the Operation Manual for the test fixture.
13. How to carry out port extension compensation (fixture selection)
Step 1. Press the
FIXTURE softkey menu.
Chapter 6
Calibration and Compensation Procedures Using Dedicated Test Fixture
button in the calibration window and measure the low-loss
LOW-LOSS C
Done
Select Fixture
key in the Cal/Compen softkey menu to display the SELECT
Calibration and Compensation
Auto Recall
check box at the time of calibration.
button in the calibration window.
127