Executing Load Calibration - Agilent Technologies Twelfth 4287A RF LCR Operation Manual

Table of Contents

Advertisement

Figure 3-18
How to connect low-loss capacitor
Step 3. Press the
cursor to the
Step 4. Press the
data. The bar in the Calibration progress indicator in the calibration window indicates the
progress of the low-loss capacitor calibration data measurement. Upon completion of
low-loss capacitor calibration data measurement, the screen changes to the next window
(Figure 3-19).

Executing load calibration

Figure 3-19
Load calibration window
Step 1. Turn the connector nut of the low-loss capacitor counterclockwise to remove it from the
7-mm terminal.
Step 2. As shown in Figure 3-20, turn the outside connector nut of the load standard (50 Ω)
counterclockwise to fully retract the internal connector sleeve.
Chapter 3
key or
key on the front panel the required number of times to move the
button.
Next>
key on the front panel to start measurement of low-loss capacitor calibration
Learning Basic Operations
Calibration and Compensation
63