External Source Mode Frequency Range - Agilent Technologies 8702D Installation Manual

Lightwave component analyzer
Table of Contents

Advertisement

Required
Equipment
Procedure
1 On the external source, press PRESET, CW, 10, MHz/msec, POWER LEVEL, –/←, 20,
2 Connect the equipment as shown in
3 On the analyzer, press PRESET, MEAS, INPUT PORTS, R.
4 Press SYSTEM, INSTRUMENT MODE, EXT SOURCE AUTO, MENU, CW FREQ, 10, M/µ.

2. External Source Mode Frequency Range

Perform this test to verify that the analyzer's reference channel, input R, is
capable of phase locking to an external CW signal.
External Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 83640A
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
Adapter, APC-3.5 (f) to APC-7 . . . . . . . . . . . . . . . . . . Agilent P/N 1250-1747
Adapter, APC-3.5 (m) to APC-7 . . . . . . . . . . . . . . . . . Agilent P/N 1250-1746
GHz/dB(m).
Figure 4-2. External Source Mode Frequency Range Test Setup
2. External Source Mode Frequency Range
Figure
4-2.
Performance Tests
4-5