Agilent Technologies Twelfth 4287A RF LCR Operation Manual page 269

Table of Contents

Advertisement

Measurement Data Transfer Time through GPIB (Typical)
Conditions:
4287A: S/N JP1KGxxxxx
Host computer: HP9000 Series, Model 715
Display: OFF
Measurement mode: List measurement
Measurement parameters: Ls and Q
Evaluation method: Average value of 10,000 repeated measurements
Table 12-4
Measurement Data Transfer Time through GPIB (Typical)
Number of measurement
Switching Time for Setup Change by GPIB (Typical)
Conditions:
Host computer: HP9000 Series, Model 715
Display: OFF
Measurement mode: List measurement
Measurement parameters: Ls and Q
Evaluation method: Average value of 10,000 repeated measurements
Table 12-5
Switching Time for Setup Change by GPIB (Typical)
measurement (required time for executing
:SOUR:LIST:TABL 1 command and
Chapter 12
Specifications and Supplemental Performance Characteristics
points
1
2
3
Conditions
Measurement table switching at list
*OPC? command)
Basic Measurement Characteristics
Required time for FETCH? command
Rdc measurement OFF
10.0 ms
12.0 ms
15.0 ms
Rdc measurement ON
10.0 ms
13.5 ms
15.0 ms
Time
8.8 ms