Setting BIN Range

Set the actual BIN range after setting the BIN range setting conditions according to
"Setting Measurement Point" on page 177), "Setting Measurement Parameter" on page
179, and "Setting Limit Mode and Reference Value" on page 181.

Setting upper limit value

Follow these steps to set the BIN upper limit value.
Step 1. Click with the mouse the
changes the button to
Step 2. Click in the cell corresponding to the column of the desired sorting condition (one of
conditions 1 to 4) on the row of the upper limit value of the BIN to be set.
Step 3. Press the entry block numeric keys (
to be set and then press the desired unit key (
To set the upper limit value to 51 nH when the limit mode is in absolute mode, for example,
press
To set the upper limit value to +1% of the reference value (Nominal) when the limit mode
is in percentage mode, press
Figure 8-9
Setting upper limit value of BIN 1
Chapter 8
button under the BIN to be set. This turns on the BIN and
OFF
ON.
[5]
,
[1],
and then
[G/n]
.
[1]
and then
Sorting According to Measurement Results
to
) to enter the upper limit value of the BIN
[0]
[9][-][.]
,
,
, or
[G/n]
[M/μ]
[k/m]
[×1]
.(Figure 8-9)