Specifications and Supplemental Performance Characteristics
Basic Measurement Characteristics
At the following points, measurement error may exceed the specifications described here
due to the 4287A's spurious characteristics:
109.7 MHz, 153.6 MHz, 177.2 MHz, 256.0 MHz, 329.1 MHz, 460.8 MHz, 768.0 MHz
Examples of Calculated Impedance Measurement Accuracy
Oscillator Level ≥ −33 dBm, Averaging Factor ≥ 8, Temperature Deviation ≤ 5°C
Figure 12-1
258
Oscillator level < −33
dBm
± (100 + 0.15 × F) [μS] (F: Frequency [MHz])
Chapter 12