14. How to display the compensation dialog box
Press the
Compensation Dialog box.
Figure 6-15
Compensation Dialog box
NOTE
The following points require attention when performing compensation data measurement.
When you start the compensation data measurement, a bar appears in the
Compensation progress indicator box of the dialog box to show how the measurement
is progressing. Upon completion of data measurement, a blue mark () appears to the
right of the executed compensation button.
When the
button is pressed, the value of the DC resistance (Rdc) measured as the compensation
data is shown in the lower-right area of the dialog box upon completion of the
respective measurement. Checking this indication at the time of compensation data
measurement makes it possible to determine that the standard is properly attached.
To erase all compensation data after measurement is finished, press the
Compensation Data
To close the dialog box during compensation, press the
15. How to select the table range for executing the compensation
Select the table range for executing the compensation by making the appropriate setting in
the
Select Compensation Table
If you select
point table (active table) selected in "2. How to select the measurement point table" on
page 135.
If you select
tables.
NOTE
If you select
tables) in the Compensation Dialog box, calibration data measurement will be performed
in accordance with the conditions set in "4. How to select the table range for executing the
calibration" on page 137.
Chapter 6
Calibration and Compensation Procedure without Using Dedicated Test
COMPEN DIALOG
key in the Cal/Compen softkey menu to display the
,
, or
OPEN
SHORT
LOAD
button.
area of the Compensation Dialog box window.
, compensation will only be executed for the measurement
Active table only
, compensation will be executed for all measurement point
Calibrate all tables
Calibrate all tables
(to execute compensation for all measurement point
Calibration and Compensation
button or the
,
DC OPEN
DC SHORT
Close
button.
Fixture
, or
DC LOAD
Clear
139