Calibration and Compensation
How to Define Compensation Kit
Figure 6-20
Compensation kit setup window
Step 2. Specify the table number to define a compensation kit by following the description in
"How to specify table number for setup" on page 104.
Step 3. Select the type of compensation kit (
selection area in the compensation kit setup window.
If you have selected "
be used as the defined compensation kit. No further setting is required.
If you have selected "
Step 4. If you want to use the same definition for all measurement points in the measurement point
table, click the
If you have selected "
point number 1. For other measurement points, the same definition as that of measurement
point number 1 is automatically entered.
Step 5. Enter open/short definitions for each measurement point.
Step 6. Enter the DC open/short definitions.
148