How to Set the Measurement Point

Unlike measuring instruments that allow constant changing of frequency and oscillator
levels during measurement, the 4287A requires that all measurement points for subsequent
measurements be set before measurement starts. The measurement point is set on the
measurement point setup display.

How to bring up the measurement point setup display

You can bring up the measurement point setup display by following this step:
Step 1. Press the
Figure 5-1
Measurement point setup display
NOTE
When you display the measurement point setup display from the setup state for the first
time, preset data has already been input for measurement point number 1 of each table (1 to
8), i.e., frequency of 1 MHz, averaging factor of 1, and oscillator level of -13.0 dBm.
When a dialog box (window for making settings) is shown, the setup display is not
switched even if the
the front panel to close the dialog box before pressing the
Chapter 5
key until the measurement point setup display appears.
[Setup View]
key is pressed. In this case, press the
[Setup View]
Setting Measurement Conditions
How to Set the Measurement Point
[Cancel]
key.
[Setup View]
key on
103