Agilent Technologies Twelfth 4287A RF LCR Operation Manual page 286

Table of Contents

Advertisement

Manual Changes
Manual Changes
screen, immediately press
NOTE
Watch the message carefully because after several seconds it will be replaced with the next
screen. Note that this is automatically done with no key operations.
If the above message does not appear, the 4287A is at fault; contact your local Agilent
customer center listed at the end of this manual or the distributor from which you
purchased the instrument.
Step 5. When "Recover Hard disk (C drive) [Y, N]?" is displayed, press
keyboard. If you want to quit the system recovery, press
Step 6. The following message appears. This is the final confirmation message asking whether you
want to start the system recover. Press
you want to quit the system recovery, press
===============
SYSTEM RECOVERY
===============
This process will recover the system drive (C:) of this
instrument to the factory-shipment state.
Please refer to the Operation Manual for more information.
This is the last chance to quit the recovery process
Continue [Y,N]?
Step 7. The system recovery will complete in about 10 minutes. The following message is
displayed during the system recovery.
===============================
SYSTEM RECOVERY IN PROGRESS....
===============================
System recovery in progress.
Please DO NOT TURN THE POWER OFF DURING THIS TIME.
CAUTION
Never turn off the power during the system recovery because doing so may cause
serious damage to the 4287A.
Step 8. When the system recovery is completed, the following message appears below the above
message indicating that the system recovery is in progress.
"Recovery Completed !"
"Please any key and then restart system."
286
on the keyboard.
on the keyboard to start the system recovery. If
to start up the 4287A as usual.
It takes about 10 minutes.
It takes about 10 minutes.
on the
to start up the 4287A as usual.
Appendix A

Advertisement

Table of Contents
loading

Table of Contents