Agilent Technologies Twelfth 4287A RF LCR Operation Manual page 60

Table of Contents

Advertisement

Learning Basic Operations
Calibration and Compensation
Step 7. Press the
cursor to the
Step 8. Press the
appears.
Figure 3-12
Window for short calibration
Step 9. The outside of the 7-mm connector of the test head installed on the test fixture stand should
be turned clockwise until the connector sleeve protrudes to its full extent at the top (Figure
3-13).
Figure 3-13
How to extend 7-mm connector sleeve
Step 10. While pressing on the short standard (0 Ω) from the top, turn it clockwise and securely
connect it to the 7-mm terminal (Figure 3-14).
60
key or
key on the front panel the required number of times to move the
button.
Next>
key on the front panel. The window for short calibration (Figure 3-12)
Chapter 3