Agilent Technologies Twelfth 4287A RF LCR Operation Manual page 123

Table of Contents

Advertisement

Calibration and Compensation Procedures Using Dedicated Test Fixture
Figure 6-5
Displaying measured DC resistance value (at calibration data measurement)
To erase all calibration data after measurement is finished, press the Clear Calibration
Data button.
To close the dialog box during calibration, press the
After turning all of the connector nuts on the standard to secure them to the 7-mm
terminals, it is not necessary to turn the connector nut of the 7-mm terminal of the test
head to secure it.
4. How to select the table range for executing the calibration
Select the table range for executing the calibration by making the appropriate setting in the
Select Calibration Table
If you select
point table (active table) selected according to "How to select the measurement point table"
on page 115.
When you select
point tables.
5. How to measure open calibration data
Step 1. With the test head mounted on the test fixture stand, turn the outer connector nut of the
7-mm connector clockwise as viewed from the top (Figure 6-6) until the connector sleeve
is fully extended.
Chapter 6
area of the Calibration Dialog box window.
, calibration will be executed only for the measurement
Active table only
, calibration will be executed for all measurement
Calibrate All Tables
Calibration and Compensation
Close
button.
123