6
Additional standards needed
Before the completion of all of the data measurement required for calculating the
calibration/compensation coefficients, a GPIB command requiring any of the measurement
data was sent. For example, when only the OPEN and SHORT measurements of the
calibration kit have been complete, the :CORR1:COLL:SAVE command, which performs
calculation of the calibration coefficients, was sent.
Measure all the required data.
-168
Block data not allowed
A block data element was received where the 4287A does not accept any block data
element.
7
Calibration required
Though the calibration was not turned on, a GPIB command was sent that is valid only
when the calibration is turned on. For example, the:CORR2:COLL command, which
obtains the data for calculating the compensation coefficients, was executed.
Turn on the calibration.
62
Can't execute data examination
The data for statistical analysis has been acquired, but change was made to the settings
during the data acquisition, so statistical analysis cannot be executed.
Acquire the data for statistical analysis with the current settings.
-148
Character data not allowed
A character data element (that did not violate the standard) was received where the 4287A
does not accept any character data element.
-100
Command error
A comprehensive syntax error occurred for which the 4287A could not detect further
details. This error code simply indicates the occurrence of a command error that is defined
in IEEE488.2,11.5.1.1.4.
11
Compensation required
Although compensation was not turned on, a GPIB command that is valid only when
compensation is turned on was sent. For example, the:DATA:CMP{1-3} command, which
reads out the compensation coefficients, was executed when compensation was turned off.
Turn on the compensation.