Specifications and Supplemental Performance Characteristics
Frequency > 500 MHz,
≤ 1 GHz
Frequency > 1 GHz,
≤ 1.8 GHz
Frequency > 1.8 GHz,
≤ 3 GHz
Oscillator level < −33
dBm
Frequency ≥ 1 MHz,
≤ 100 MHz
Frequency > 100 MHz,
≤ 500 MHz
Frequency > 500 MHz,
≤ 1 GHz
Frequency > 1 GHz,
≤ 1.8 GHz
Frequency > 1.8 GHz,
≤ 3 GHz
Eb =
Ec =
Zs =
Oscillator level ≥ −33
dBm, Averaging factor ≥ 8
Oscillator level ≥ −33
dBm, Averaging factor < 7
Oscillator level < −33
dBm
Yo =
Oscillator level ≥ −33
dBm, Averaging factor ≥ 8
Oscillator level ≥ −33
dBm, Averaging factor < 7
Chapter 12
Basic Measurement Characteristics
±1.2 [%]
±2.5 [%]
±5 [%]
±1 [%]
±1.2 [%]
±1.2 [%]
±2.5 [%]
±5 [%]
Z
s
±
×
------- -
+
Y
Z
100
[%]
o
x
Z
x
(|Zx|: Measurement value of |Z|)
×
0.08
F
±
------------------- -
0.06
+
[%] (F: Frequency [MHz])
1000
Within ±5°C from the calibration temperature.
Measurement accuracy applies when the calibration is
performed at 23°C ±5°C. When the calibration is performed
beyond 23°C ±5°C, the measurement accuracy decreases to
half that described.
± (20 + 0.5 × F) [mΩ] (F: Frequency [MHz])
± (50 + 0.5 × F) [mΩ] (F: Frequency [MHz])
± (100 + 0.5 × F) [mΩ] (F: Frequency [MHz])
Within ±5°C from the calibration temperature.
Measurement accuracy applies when the calibration is
performed at 23°C ±5°C. When the calibration is performed
beyond 23°C ±5°C, the measurement accuracy decreases to
half that described.
± (30 + 0.15 × F) [μS] (F: Frequency [MHz])
± (50 + 0.15 × F) [μS] (F: Frequency [MHz])
257